Sign In | Join Free | My gimpguru.org
China Guangzhou UC Instruments., Co. Ltd. logo
Guangzhou UC Instruments., Co. Ltd.
Guangzhou UC Instruments., Co. Ltd.
Verified Supplier

4 Years

Home > Optical Oscilloscope >

1600nm To 800nm Full Rate Optical Oscilloscope 10Gbps CDR

Guangzhou UC Instruments., Co. Ltd.
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now

1600nm To 800nm Full Rate Optical Oscilloscope 10Gbps CDR

Brand Name : YOUCii

Model Number : U9710A

Place of Origin : CHINA

MOQ : 1pcs

Price : negotiate a price

Payment Terms : L/C, D/A, D/P, T/T, Western Union, MoneyGram

Supply Ability : 5 pcs+per month

Delivery Time : 20 work days

Packaging Details : Carton

Certification : CU,IOS

Product name : Full Rate Optical Oscilloscope

Wavelength range : 1600nm To 800nm

Data rate range : 0.155 - 11.3 Gbps

Operating temperature : -40 ~ 85℃

Model : U9710A

Brand : YOUCII

Contact Now

U9710A Full Rate 10Gbps CDR

No. Specifications Max. Min. Unit.
1 Data rate range 11.3 0.155 Gbps
2 Wavelength range 1600 800 nm
3 Optical input range (determined by module sensitivity) +3 -15 dBm
4 Output recovered electrical clock range 5.65 / 2.825 0.0775/0.03875 GHz
5 Output electrical clock amplitude 600(typical value) mVpp
6 Clock rise time(OC-192)(20%~80%) 33.1 22.2 ps
7 Clock fall time(OC-192)(80%~20%) 33.7 23.9 ps
8 Output Jitter(10.3125Gbps) 1 (typical value) ps
9 Clock frequency(10.3125Gbpsrate, divided by2/4) 5.156 / 2.578 5.156 / 2.578 GHz
10 Internal Loss of Signal Detect (LOS) 128 (overload) 5 (sensitivity) mV
11 Power supply voltage (input voltage of external power supply) 5.5 4 V
12 Operating temperature 85 -40

U9710 CDR (Clock Recovery Module ) is mainly used to provide clock recovery signal in the test environment without triggering clock, so as to use the sampling oscilloscope to test the optical eye diagram of the device. The typical application is the testing of optical network equipment or the BOB product in PON application. Because the photometric device itself has no frequency division clock output, it is necessary to recover the clock signal to trigger the sampling oscilloscope when testing the optical eye diagram.
U9710 CDR has a SFP optical port, which is suitable for SFP encapsulated conventional SFP, SFP+, OLT SFP and other optical modules. It only needs the receiving end of the optical module to convert the optical signal to the electrical signal, and then extract and output the clock of the electrical signal. Clock signal is mainly used for the trigger signal of oscilloscope. This device self-adapts to the rate and does not require software configuration.

FAQ
Q: Are your company a trading one or a factory?
A: We are a fcatory in Guangdong. Our company has a professional foreign trade sales team and an experienced technical support team. We are supplier for Light source, optical power meter, optical attenuator, optical switch and BERT, DCA, CDR.
Q: What is the price term?
A: FOB ShenZhen
Q: What’s the warranty time?
A: 12 months after shipment date.
Q: How to place order?
A: Talk to service staff, confirm machine details, and then place order.


Product Tags:

Optical Oscilloscope 10Gbps CDR

      

1600nm optical sampling oscilloscope

      

800nm Optical Oscilloscope

      
Quality 1600nm To 800nm Full Rate Optical Oscilloscope 10Gbps CDR wholesale

1600nm To 800nm Full Rate Optical Oscilloscope 10Gbps CDR Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: Guangzhou UC Instruments., Co. Ltd.
*Subject:
*Message:
Characters Remaining: (0/3000)